FDAClass IISeptember 22, 2021

Medtronic HVAD Pump Implant Kits recall

Medtronic HVAD Pump Implant Kits

Why was Medtronic HVAD Pump Implant Kits recalled?

There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than over an unrepaired strain relief. This may lead to difficulty in pulling back the driveline cover when attempting to access the Driveline connector during a controller exchange. Additionally, sliding the driveline cover over a strain relief repair may compromise the repair.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product Numbers: 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU - all serial numbers

Product
Medtronic HVAD Pump Implant Kits
Recalling firm
Heartware, Inc.
Quantity
11851 (potentially active, US); 1995 (active, US)
Distribution
Distribution throughout the US. Also distribution in ARGENTINA Armenia, Australia, Austria, BELGIUM, BRAZIL, CANADA, CHILE, Colombia, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, Germany, GREECE HONG KONG, HUNGARY, INDIA, ISRAEL, Italy, JAPAN, Kazakhstan, Korea, KUWAIT, Latvia, LEBANON, LITHUANIA, LUXEMBOURG, Macedonia, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, Poland, ROMANIA, SAUDI ARABIA, Serbia, Singapore, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, Taiwan, TURKEY, Ukraine, United Arab Emirates, United Kingdom, VIETNAM
Recall date
September 22, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls