Lyra Direct SARS-CoV-2 Assay Emergency Use recall
Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization- Ref: M124. The Lyra Direct SARS-CoV-2 Assay is a real-time RT-PCR assay intended for the…
Why was Lyra Direct SARS-CoV-2 Assay Emergency Use recalled?
Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle-threshold (Ct) value of 5 when using the Thermo Fisher QuantStudio 7 Pro, the Applied Biosystems 7500 Fast Dx, the Applied Biosystems 7500, the Bio-Rad CFX96 Touch", the Roche LightCycler LC 480 II/ cobas z 480, or the Qiagen Rotor-Gene¿ MDx.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lots
- Product
- Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization- Ref: M124. The Lyra Direct SARS-CoV-2 Assay is a real-time RT-PCR assay intended for the…
- Recalling firm
- Quidel Corporation
- Quantity
- 81,910 kits
- Distribution
- Worldwide distribution - US Nationwide distribution including Puerto Rico and the countries of Canada, Panama, Virgin Islands, Germany, Honduras.
- Recall date
- September 22, 2021
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)