FDAClass IIJuly 25, 2018

t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM recall

t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.

Why was t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM recalled?

The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways: A) by triggering a Malfunction 4 Alarm, or B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alarm 12 (Battery Very Low). The alarms in both scenarios notify the user that the pump has stopped delivering insulin.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 483731, 483930, 483992, 484013, 492055, 492368, 501157, 501177, 501178, 501187, 501191, 501200, 501202, 501205, 501207, 501268, 501285, 501364, 501368, 501370, 501376, 501385, 501386, 501387, 501432, 501435, 501440, 501443, 501444, 501445, 501466, 501479, 501488, 501490, 501491, 501495, 501503, 501508, 501509, 501519, 501522, 501523, 501526, 501535, 501560, 504522, 504531, 504535, 504540, 504753, 504972, 504990, 505025, 90472671, 90488584

Product
t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
Recalling firm
Tandem Diabetes Care Inc
Quantity
55
Distribution
U.S. Consignees: IN, AZ, UT, TX, NY, NJ, MD, MN, IL, CO, AL, WI, CA, PA, OH, OR, MS, FL, NM, ID, NV, KY, VA,
Recall date
July 25, 2018
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls