FDAClass IIJuly 8, 2020

ABL80-FLEX-393-839 Box Label recall

ABL80-FLEX-393-839 Box Label, REF 393-839, IVD, CE, UDI: 05700693938394 - Product Usage: It is intended for use in a laboratory environment, near…

Why was ABL80-FLEX-393-839 Box Label recalled?

Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All lots

Product
ABL80-FLEX-393-839 Box Label, REF 393-839, IVD, CE, UDI: 05700693938394 - Product Usage: It is intended for use in a laboratory environment, near…
Recalling firm
Radiometer Medical ApS
Quantity
9373 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AK, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CA.
Recall date
July 8, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls