Breathing circuit set recall
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Why was Breathing circuit set recalled?
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
REF/UDI-DI: 260127/07630002802956, 17630002802953; 260128/07630002802963, 17630002802960; 260167/07630002802970, 17630002802977; 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)
- Product
- Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
- Recalling firm
- Hamilton Medical AG
- Quantity
- 27,270 pcs
- Distribution
- US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.
- Recall date
- July 8, 2026
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)