FDAClass IISeptember 2, 2015

Central Line >34 Weeks Kit. Cardiology department. The recall

Central Line >34 Weeks Kit. Cardiology department. The Central Line >34 Weeks Kit (DYNDC1987A) is intended for use on patients greater than 34 weeks…

Why was Central Line >34 Weeks Kit. Cardiology department. The recalled?

The print plate artwork for this lot was found to state: Central Line <34 Weeks Kit displaying an incorrect sign in front of the 34. The description for this device should have stated Central Line >34 Weeks Kit. The defect may lead to the product to be used on patient population that is <34 weeks rather than intended population of >34 weeks.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Central Venous Catheter Kit for patients >34 weeks old with the coding below were affected. Lot Numbers: 14DB9277 and 14PB2878 Unit No: DYNDC1987A Expiration Date: 06/2015 Serial No: N/A UPC code: N/A

Product
Central Line >34 Weeks Kit. Cardiology department. The Central Line >34 Weeks Kit (DYNDC1987A) is intended for use on patients greater than 34 weeks…
Recalling firm
Medline Industries Inc
Quantity
320 units for lot 14DB9277 & 320 units for lot 14PB2878
Distribution
US including PA
Recall date
September 2, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls