FDAClass IIAugust 7, 2024

Arctic Sun Temperature Management Systems recall: Software Issue

Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU…

Why was Arctic Sun Temperature Management Systems recalled?

If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF/UDI-DI/Software: 6000-00-00/00801741161513/v1.0.5 and prior, 6000-00-00L/00801741161520/v1.0.5 and prior, 600001/00801741226380/v2.1.1.0 and prior, 600006/00801741226397/v2.1.1.0 and prior, 600020/00801741226403/v2.1.1.0 and prior, 5000-00-00/00801741080142/v3.0.2 and prior, 5000-00-00E/00801741127755/v3.0.2 and prior, 5000-00-00L/00801741080159/v3.0.2 and prior, 5000-1-01/00801741186134/v3.0.2 and prior, 5000-01-01L/00801741186141/v3.0.2 and prior, 5000-01-02/00801741170003/v3.0.2 and prior, 5000-01-03/00801741222818/v3.0.2 and prior, 5000-01-04/00801741144653/v4.0.0 and prior, 5000-01-05/00801741222825/v3.0.2 and prior, 5000-01-07/00801741222832/v3.0.2 and prior, 5000-01-08/00801741222849/v3.0.2 and prior All serial numbers.

Product
Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU…
Recalling firm
Medivance Inc.
Quantity
10507
Distribution
Worldwide - US Nationwide distribution including in the states of IN, OK, MO, FL, CT, KS, TX, NJ, CA, OR, HI, IL, WI, VA, KY, NH, CO, NV, LA, NC, ID, UT, GA, SC, TN, MD, PA, IA, MA, OH, NY, DE, MI, NM, WV, AZ, DC, VT, MN, SD, AL, AR, WA, NE, ND, MT, AK, MS, ME, WY, RI, PR, OK, OR and the countries of CA, DE, AT, BE, FI, TR, NO, DK, JP, ES, AU, FR, GB, IT, SE, PT, AE, SA, ZA, RU, KR, CY, TH, GR, TW, MY, OM, CN, SG, GI, NL, IL, CZ, RO, CH, RE, PL, PE, SK, AR, IN, CL, MX, QA, BR, SI, HU, JO, LU, KW, PH, NZ, HK, VN, IR, BN, UA, ID, CO, KZ, HR, IS, GU, GE, BH, EE, AM, PA, KR, CL, TH, MY, JP, AU, BD, BE, TW
Recall date
August 7, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)