Micro Joystick R-net. Electrical wheelchair component recall: Software Issue
Micro Joystick R-net. Electrical wheelchair component.
Why was Micro Joystick R-net. Electrical wheelchair component recalled?
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.
- Product
- Micro Joystick R-net. Electrical wheelchair component.
- Recalling firm
- mo-Vis BVBA
- Quantity
- 3 units
- Distribution
- US Nationwide distribution in the state of TX.
- Recall date
- September 10, 2025
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)