FDAClass IISeptember 2, 2015

CARESTREAM DRX-EVOLUTION X-Ray System Product Usage recall: Software Issue

CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray…

Why was CARESTREAM DRX-EVOLUTION X-Ray System Product Usage recalled?

Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software versions 5.6 and greater and 5.7 (current releases); Serial Numbers: 5029, 5569, 4067, 4754, 4131, 4100, 4096, 4076, 4569, 4570, 4588, 4670, 4767, 4879, 5266, 5393, 5408, 5442, 5572, 5573, 5555, 5556, 5563, 4146, 4679, 4702, 4703, 4704, 4801, 5090, 5154, 4787, 4069, 4074, 4344, 4085, 4058, 4084, 5470, 5158, 4086, 4099, 4073, 4665, 5241, 5303, 5304, 4639, 5350, 5352, 5429, 5508, 4696, 4763, 5182, 4291, 5037, 5554, 4192, 4117, 4172, 4207, 4225, 4239, 4278, 4301, 4310, 4328, 4333, 4338, 4366, 4384, 4421, 4427, 4442, 4443, 4447, 4465, 4499, 4523, 4534, 4538, 4539, 4692, 4822, 4861, 4875, 5128, 5131, 5174, 5553, 5579, 5580, 5597, 5628, 4320, 5309, 5042, 4807, 4808, 4809, 4810, 5493, 4266, 4711, 5327, 4125, 4143, 4097, 4156, 4282, 4287, 4289, 4258, 4742, 4964, 4784, 5050, 5551, 4233, 4644, 4082, 4113, 4196, 4683, 4309, 4315, 4064, 4256, 4987, 4624, 4585, 4882, 5362, 4004, 5377, 5418, 4300, 4308, 4627, 4638, 5138, 5251, 5360, 4947, 5620, 4985, 5403, 5495, 5560, 5544, 5545, 5621, 5622, 5451, 5282, 5645, 4724, 4298, 4269, 4756, 4418, 5273, 5178, 5179, 5443, 5197, 5389, 5584, 5473, 4899, 5334, 5431, 4990 --- NOTE: Additional serial numbers at foreign consignees to be determined upon reporting part usage.

Product
CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray…
Recalling firm
Carestream Health Inc
Quantity
Domestic: 76 units, Foreign: 160 units
Distribution
Worldwide Distribution - US Nationwide and in the countries: Australia, Canada, Chile, China, India, Korea, Malaysia, New Zealand, Singapore, Switzerland, Taiwan and Thailand.
Recall date
September 2, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls