FDAClass IIAugust 30, 2023

Randox Liquid Urine Control Level 2- intended for in recall

Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol, Creatinine…

Why was Randox Liquid Urine Control Level 2- intended for in recalled?

(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN: 05055273207569 Batch/Lot Number: 1209UC Expiry Date: 28 Mar 24

Product
Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol, Creatinine…
Recalling firm
Randox Laboratories Ltd.
Quantity
39 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, FL, GA, OH, VA including PR and WW distribution.
Recall date
August 30, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)