PTW UNIDOS webline: 1) recall
PTW UNIDOS webline: 1) REF T10023, 2) REF T10022 and 3) REF T10023 Firmware version 2.05. Dosimeter for radiation therapy, diagnostic radiology and…
Why was PTW UNIDOS webline: 1) recalled?
In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. This could result in a radiation over-exposure.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
serial numbers: **T10021: 000005, 000019, 000037, 000041, 000047, 000049, 000058, 000066, 000074, 000076, 000079, 000099, 000107, 000125, 000141, 000144, 000158, 000161, 000168, 000175, 000186, 000212, 000221, 000224, 000229, 000231, 000254, 000262, 000274, 000294, 000301, 000302, 000327, 000351, 000352, 000382, 000401, 000413, 000424, 000437, 000445, 000452, 000470, 000477, 000486, 000504, 000525, 000528, 000535, 000567, 000593, 000594, 000649, 000655, 000658, 000659, 000663, 000678, 000686, 000710, 000711, 000731, 000752, 000756, 000763, 000785, 000792, 000793, 000796, 000800, 000820, 000829, 000830, 000857, 000860, 000863, 000865 ,000871, 000874, 000890, 000892, 000946, **T10022: 000003, 000046, 000066, 000087, 000096, 000106, 000118, 000138, 000172, 000178, 000183, 000186, 000188, 000210, 000240, 000317, 000336, 000347, 000350, 000351, 000358, 000360, 000392, 000398, 000420, 000422, 000425, 000434 -000453, 000456-000469, 000474, ** T10023: 000022, 000028, 000051, 00102, 000196, 000213, 000226, 000250, 000255, 000257, 000259, 000261, 000263, 000271, 000280, 000294
- Product
- PTW UNIDOS webline: 1) REF T10023, 2) REF T10022 and 3) REF T10023 Firmware version 2.05. Dosimeter for radiation therapy, diagnostic radiology and…
- Recalling firm
- PTW-Freiburg
- Quantity
- 242 firmwares
- Distribution
- Distributed in CA, MI, MA, NC, NJ, NY, PA, TN, TX, PA & WI.
- Recall date
- September 2, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)