FDAClass ISeptember 22, 2021

Cordis Super Torque MB 5f PIG recall

Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598C

Why was Cordis Super Torque MB 5f PIG recalled?

Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

all codes

Product
Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598C
Recalling firm
Cordis Corporation
Quantity
26025 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Europe.
Recall date
September 22, 2021
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls