FDAClass ISeptember 10, 2025

WATCHMAN TruSeal Access System ANT recall

WATCHMAN TruSeal Access System ANT, OUS, Material Number (UPN) M635TS70040; intended to provide vascular and transseptal access for all WATCHMAN Left…

Why was WATCHMAN TruSeal Access System ANT recalled?

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN 08714729965756, ALL NON-EXPIRED BATCHES

Product
WATCHMAN TruSeal Access System ANT, OUS, Material Number (UPN) M635TS70040; intended to provide vascular and transseptal access for all WATCHMAN Left…
Recalling firm
Boston Scientific Corporation
Quantity
340185 units in total
Distribution
Worldwide.
Recall date
September 10, 2025
Classification
Class I
Status
Ongoing
Agency
FDA (device)