FDAClass IIJuly 18, 2018

Medtronic EnTrust SingleChamber Implantable recall

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product…

Why was Medtronic EnTrust SingleChamber Implantable recalled?

Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product UPN/GTIN: 00885074300245 Serial Numbers: PNT402938H PNT413911H PNT413995H

Product
Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product…
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
3 units
Distribution
Worldwide Distribution - US Nationwide
Recall date
July 18, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)