FDAClass IIAugust 21, 2024

Orthopantomograph OP300 dental computed tomography recall

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for…

Why was Orthopantomograph OP300 dental computed tomography recalled?

It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

DI: 06430035872725, 06430035872732, 06430035872749, 06430035872756, 06430035872763, 06430035872770

Product
Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for…
Recalling firm
Palodex Group Oy
Quantity
103
Distribution
US Nationwide-Worldwide
Recall date
August 21, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls