Surgical Image Guidance Functionality recall: Software Issue
Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
Why was Surgical Image Guidance Functionality recalled?
After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: 00380650003997. Software Version: 1.5. Serial Numbers: 004-0680001, 004-0680002, 004-0680003, 004-0680004, 004-0680007, 004-0680008, 004-0680017, 004-0680026, 004-0680036, 004-0680047, 004-0680048, 004-0680086, 004-0680090, 004-0680092, 004-0680096, 004-0680100, 004-0680107, 004-0680112, 004-0680120, ,004-0680127,004-0680131, 004-0680134, 004-0680140, 004-0680152, 004-0680153, 004-0680161, 004-0680163, 004-0680188, 004-0680190,004-0680203, 004-0680206, 004-0680266, 004-0680267, 004-0680271, 004-0680273, 004-0680274, 004-0680283, 004-0680288, 004-0680293, 004-0680294, 004-0680306, 004-0680311, 004-0680315, 004-0680319, 004-0680325, 004-0680334, 004-0680336, 004-0680337, 004-0680338, 004-0680339, 004-0680346, 004-0680348, 004-0680352, 004-0680362, 004-0680404, 004-0680417, 004-0680418, 005-7230003, 005-7230005, 005-7230007, 005-7230017, 005-7230019, 005-7230021, 005-7230022, 005-7230023, 005-7230031, 005-7230035, 005-7230057, 005-7230081, 005-7230085, 005-7230086, 005-7230102, 005-7230108, 005-7230109, 005-7230186, 007-4610018, 007-4610023, 007-4610030, 007-4610032, 007-4610033, 007-4610034, 007-4610037, 007-4610044, 007-4610074, 007-4610107, 007-4610112, 007-4610115, 007-4610122, 007-4610123, 007-4610124, 007-4610125, 007-4610127, 007-4610132, 007-4610138, 007-4610166, 007-4610167, 007-4610179, 007-4610181, 007-4610186, 007-4610217, 007-4610219, 007-4610220, 007-4610226, 007-4610229, 007-4610230, 007-4610262, 007-4610275, 007-4610283, 007-4610287, 007-4610288, 007-4610308, 007-4610309, 007-4610351, 007-4610355, 007-4610359, 007-4610360, 007-4610366, 007-4610394, 007-4610404, 007-4610405, 007-4610406, 007-4610409, 009-4570001, 009-4570004, 009-4570005, 009-4570007, 009-4570013, 009-4570014, 009-4570015, 009-4570016, 009-4570019, 009-4570022, 009-4570033, 009-4570035, 009-4570044, 009-4570047, 009-4570048, 009-4570049, 009-4570051,009-4570052, 009-4570053, , 009-4570054,009-4570057, 009-4570059, 009-4570060, 009-4570064, 009-4570068, 009-4570070, 009-4570072, 009-4570073, 009-4570096, 009-4570097, 009-4570099, 009-4570100, 009-4570114, 009-4570116, 009-4570125, 009-4570127, 009-4570129, 009-4570130, 009-4570136, 009-4570137, 009-4570144, 009-4570148, 009-4570149, 009-4570150, 009-4570153, 009-4570155, 009-4570156, 009-4570158, 009-4570160, 009-4570161, 009-4570167, 009-4570171, 009-4570173, 009-4570176,009-4570178, 009-4570179, 009-4570180, 009-4570183, 009-4570184, 009-4570200, 009-4570204, 009-4570207, 009-4570233, 009-4570249, 009-4570256, 009-4570259, 009-4570263, 009-4570270, 009-4570275, 009-4570278, 009-4570281, 009-4570289, 009-4570293, 009-4570294, 009-4570300, 009-4570308, 009-4570317, 009-4570320, 009-4570331, 009-4570342, 009-4570343, 009-4570350, 009-4570353, 009-4570357, 009-4570365, 009-4570368, 009-4570393, 009-4570397, 009-4570425, 009-4570431, 009-4570435, 009-4570440, 009-4570448, 009-4570449, 009-4570451, 009-4570455, 009-4570457, 009-4570459, 009-4570460, 009-4570474, 009-4570489, 009-4570508, 009-4570510, DM550021, DM550024, DM550043, DM550051, DM550054
- Product
- Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
- Recalling firm
- Alcon Research LLC
- Quantity
- 230
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA, WI, PR and the countries of Canada, Japan, France.
- Recall date
- July 24, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)