Atellica Solution IM1600 Analyzer recall
Atellica Solution IM1600 Analyzer, Siemens Material Number 11066000, UDI # 00630414002026
Why was Atellica Solution IM1600 Analyzer recalled?
Insufficient sample may not be detected when using Tube Top Sample Cup (TTSC) or V-shaped Sample Cup sample containers with the Atellica IM1300 and Atellica IM1600 analyzers.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Atellica Tube Top Sample Cups, 1 mL, SMN 11069061; Atellica Tube Top Sample Cups, 2 mL, SMN 11069062; Sarstedt 62.612, 15.3 x 92 False Bottom, 4 mL
- Product
- Atellica Solution IM1600 Analyzer, Siemens Material Number 11066000, UDI # 00630414002026
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 1329
- Distribution
- Worldwide distribution including to the following states: Florida, New Jersey, New York, California, Arizona, Colorado, Pennsylvania, Virginia, Ohio, Texas, Alabama, Nebraska, Wisconsin, Montana, Kentucky, Washington DC, Massachusetts, North Carolina, Georgia, Louisiana, Kansas, South Carolina, Tennessee, Iowa, Rhode Island, South Dakota, New Hampshire, Maryland, Oregon, Indiana, Idaho, Connecticut, Puerto Rico, West Virginia, Washington, Illinois, Utah, Maine, Minnesota, Wyoming, Nevada, Michigan, Missouri, and New Mexico.
- Recall date
- July 1, 2020
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)