Medicine Cup recall
Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.
Why was Medicine Cup recalled?
There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers 122211502 122212515
- Product
- Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.
- Recalling firm
- Medline Industries Inc
- Quantity
- 903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups)
- Distribution
- Continental U.S.
- Recall date
- August 17, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)