FDAClass IIAugust 17, 2016

Medicine Cup recall

Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.

Why was Medicine Cup recalled?

There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers 122211502  122212515

Product
Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.
Recalling firm
Medline Industries Inc
Quantity
903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups)
Distribution
Continental U.S.
Recall date
August 17, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)