FDAClass IISeptember 8, 2021

TRUMATCH CT FEM CUT TIB PIN R- TruMatch Patient recall: Nitrosamine Impurity

TRUMATCH CT FEM CUT TIB PIN R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in…

Why was TRUMATCH CT FEM CUT TIB PIN R- TruMatch Patient recalled?

Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot number/ GTIN: 27975, 27470 GTIN: 10603295384823

Product
TRUMATCH CT FEM CUT TIB PIN R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in…
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
2 cases (1/cs)
Distribution
Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK.
Recall date
September 8, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls