FDAClass IISeptember 2, 2015

Philips DigitalDiagnost Release 4.0.3 recall

Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in…

Why was Philips DigitalDiagnost Release 4.0.3 recalled?

The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7 operating system.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

DigitalDiagnost

Product
Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in…
Recalling firm
Philips Electronics North America Corporation
Quantity
62
Distribution
US Nationwide Distribution to the states of : Arizona, Minnesota, State of Washington, Wyoming and Ohio. Internationally to Canada.
Recall date
September 2, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls