FDAClass IIAugust 17, 2016

Merge PACS software. Product was recall

Merge PACS software. Product was distributed under the AMICAS label, Brighton, MA, but labeling was later changed to Merge Healthcare, Hartland, WI.

Why was Merge PACS software. Product was recalled?

The patient name in the Halo title bar and the thumbnails do not match the name on displayed images.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Versions: V6.0.2.0 MR2 and earlier

Product
Merge PACS software. Product was distributed under the AMICAS label, Brighton, MA, but labeling was later changed to Merge Healthcare, Hartland, WI.
Recalling firm
Merge Healthcare, Inc.
Quantity
93 sites have the affected version
Distribution
The software was distributed to medical facilities nationwide. Government distribution was made to MN. Foreign distribution was made to Australia and to the United Kingdom. There was no military distribution.
Recall date
August 17, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls