QIAcube Connect MDx - IVD Instrument designed to recall
QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or…
Why was QIAcube Connect MDx - IVD Instrument designed to recalled?
Heater shaker module does not perform heating if the temperature is set below 40¿C for static incubation steps when running IVD protocol QIAamp¿ DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA from Gram-positive bacteria (Script filename: Kt_name_Bacteria (Gram+) or rY'east_Enzymatic Lysis_ V2) which could potentially cause delayed or erroneous results depending on the downstream assay.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GTIN 04053228039129 ALL serial numbers shipped to date. SN 30000-30506 is the impacted range (later serial numbers and units in QIAGEN warehouses will include supplemental instructions for operating the heater in protocols requiring incubation temperatures of 40C or lower prior to shipment)
- Product
- QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or…
- Recalling firm
- Qiagen Sciences LLC
- Quantity
- 419 units
- Distribution
- Nationwide. Foreign: AD, AT, AU, BE, CA, CH, CN, CZ, DE, DK, ES, FI, FR, GB, ID, IN, IT, JO, KR, KZ, LU, ML, MY, NL, NO, PH, PL, PY, SA, SE, SG, SK, SO, VN J, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI
- Recall date
- July 24, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)