Dynex Agility Sample Tips recall
Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA…
Why was Dynex Agility Sample Tips recalled?
The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number: 67910; UDI-DI: 5060456180072; Lot Number: 863366;
- Product
- Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA…
- Recalling firm
- Dynex Technologies, Inc.
- Quantity
- 237 units
- Distribution
- Domestic: CA, FL, NC, NJ, NY;
- Recall date
- August 27, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)