The SYSTEM 1E Liquid Sterilant Processing System. The recall
The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization…
Why was The SYSTEM 1E Liquid Sterilant Processing System. The recalled?
The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model #: P6500, S/N 400000 through 406294
- Product
- The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization…
- Recalling firm
- Steris Corporation
- Quantity
- 6,051 units
- Distribution
- Worldwide Distribution -- USA (nationwide) and the countries of Bahrain and Hong Kong.
- Recall date
- September 19, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)