FDAClass IAugust 27, 2025

Reliance 4-front recall

RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658…

Why was Reliance 4-front recalled?

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model 0657 UDI-DI 00802526560934; Model 0682 UDI-DI 00802526560958; Model 0683 UDI-DI 00802526560965; Model 0685 UDI-DI 00802526560972; Model 0686 UDI-DI 00802526560989; Model 0692 UDI-DI 00802526519017 00802526560996 00802526568664 00802526568688; Model 0693 UDI-DI 00802526518850 00802526519055 00802526537769 00802526561009 00802526568701; Model 0695 UDI-DI 00802526519093 00802526561016; Model 0696 UDI-DI 00802526519987 00802526561207 00802526568831

Product
RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658…
Recalling firm
Boston Scientific Corporation
Quantity
247,702 US; 344,670 OUS
Distribution
Worldwide Distribution. US nationwide. International distribution worldwide.
Recall date
August 27, 2025
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls