Why was EPIQ Ultrasound Systems recalled?
Ultrasound system control panel arm swivel lock could fail, which could cause the control panel and monitor to swivel freely while a user is positioning the monitor, or transporting/moving the system, which could lead to patient or user musculoskeletal injury.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Manufactured Between 2020-02-27 to 2021-05-25
- Product
- EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx, EPIQ CVxi
- Recalling firm
- Philips Ultrasound Inc
- Quantity
- 8935
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of FL, UT, CA, CO, WA, MS, NC, NV, GA, LA, VA, SC, NE, TN, HI, OR, ID, MI, OH, IN, IL, OK, TX, PA, NJ, CT, NY, DE, SD, WI, VT, MN, KS, MO, KY, MA, NH, IA, MT, NM, ME, AZ, AR, AL, WY, ND, AK, MD, RI, WV, DC, VI and the countries of Argentina, Australia, Bahrain, Peru, Serbia, Bolivia, Moldova, Albania, Ukraine, Uruguay, Belarus, Hungary, Greece, Honduras, Montenegro, Senegal, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Dominican Republic, Germany, Ecuador, Estonia, Finland, Taiwan, Georgia, United Kingdom, Guatemala, Hong Kong, Indonesia, Iran, Israel, Yemen, Jordan, Korea, Republic of, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Morocco, Mauritius, Nicaragua, New Zealand, Norway, Oman, United Arab Emirates, Austria, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Thailand, Czech Republic, Tunisia, Uzbekistan, South Africa, Sweden, Switzerland, Netherlands, France, Monaco, Martinique, New Caledonia, Italy, China, India, Denmark, Singapore, Belgium, Spain, Mexico, Viet Nam, Philippines, Malaysia, Japan, Bangladesh, Ireland, Luxembourg, Andorra, Nepal, Myanmar, Pakistan, Brunei Darussalam, French Guiana, Sri Lanka, Kazakhstan, Egypt, Turkey, Kuwait, Romania, Slovakia, Croatia, Slovenia, Bosnia and Herzegovina.
- Recall date
- August 25, 2021
- Classification
- Class II
- Status
- Completed
- Agency
- FDA (device)