FDAClass IIJune 17, 2020

Elekta Unity recall

Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571

Why was Elekta Unity recalled?

In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571. UDI Number 05060191071321. Anatomy (Protocol Name): Abdomen (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Fast Sag Cor RealTime, btFFE Sag RealTime); HeadNeck L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); HeadNeck M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Thorax (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Sag RealTime, btFFE Sag RealTime)

Product
Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571
Recalling firm
Elekta Limited
Quantity
26 units
Distribution
US distribution to IA, NY, PA, TX, WI. International distribution to Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Turkey, United Kingdom.
Recall date
June 17, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls