FDAClass IIAugust 25, 2021

Trilogy EV300 recall: Software Issue

Trilogy EV300, Product numbers CA2200X12B, DS2200X11B, FX2200X15B, IN2200X15B

Why was Trilogy EV300 recalled?

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02

Product
Trilogy EV300, Product numbers CA2200X12B, DS2200X11B, FX2200X15B, IN2200X15B
Recalling firm
Philips Respironics, Inc.
Quantity
18936 (US); 4659 (OUS)
Distribution
Worldwide distribution - US nationwide distribution.
Recall date
August 25, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)