FDAClass IIAugust 21, 2019

ARIES Extraction Kit recall

ARIES Extraction Kit; ARIES System; SYNCT , REF 50-10026, UDI # 00840487100073

Why was ARIES Extraction Kit recalled?

Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot numbers & discontinue use by dates for Part Number 50-10026: AA7126 Immediately AA7307 - Immediately AA7385 - Immediately AA7525 - 3/1/2019 AA7547 - 3/16/2019 AA7628 - 4/12/2019 AA7666 - 3/28/2019

Product
ARIES Extraction Kit; ARIES System; SYNCT , REF 50-10026, UDI # 00840487100073
Recalling firm
Luminex Corporation
Quantity
N/A
Distribution
US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
Recall date
August 21, 2019
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls