FDAClass IIJuly 17, 2024

ARTHO KNEE BAPTIST SOUTH PACK Procedure tray recall

ARTHO KNEE BAPTIST SOUTH PACK Procedure tray Catalog Number: BPKA49B

Why was ARTHO KNEE BAPTIST SOUTH PACK Procedure tray recalled?

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 00191072197931 LOT#"s: 953241 882241

Product
ARTHO KNEE BAPTIST SOUTH PACK Procedure tray Catalog Number: BPKA49B
Recalling firm
American Contract Systems Inc
Quantity
129 units
Distribution
US Nationwide distribution.
Recall date
July 17, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)