Field Upgrade Kits: P/N 5451208 RIU recall
Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE…
Why was Field Upgrade Kits: P/N 5451208 RIU recalled?
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
P/N/Sales Order Number: 5451208/80165102, 80171955, 80173514, 80162453, 80183950, 80151733, 80163547, 80171532, 80156410, 80176650, 10530514, 105449333, 80167595; 5485304/10557644, 118687327; 5761613/10283014, 10152211, 105108003, 105114645. OEC 9800, UDI-DI: 00840682114349; OEC 9900 Systems, UDI-DI: 00840682114301
- Product
- Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE…
- Recalling firm
- GE OEC Medical Systems, Inc
- Quantity
- 16
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
- Recall date
- August 9, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)