Mini-Infuser recall
Mini-Infuser Model 300XL Multispeed Infusion Pump, an Rx battery operated syringe infusion pump with an active time scale and rate selector switch…
Why was Mini-Infuser recalled?
Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
product codes 2M8171 and 2M8171R, all serial numbers
- Product
- Mini-Infuser Model 300XL Multispeed Infusion Pump, an Rx battery operated syringe infusion pump with an active time scale and rate selector switch…
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 19,380 units
- Distribution
- 45,865 units
- Recall date
- September 5, 2012
- Classification
- Class III
- Status
- Terminated
- Agency
- FDA (device)