Nitinol Staple recall
Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22
Why was Nitinol Staple recalled?
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot #: a) R900, 009J, 050511, 00AG, 0095, 183601, 10505-11, 193601; b) R903, 26404, 264-04, 03NK, 050512, 10505-12, 24604, 11811, 181102, 181103, 188803; c) 26405, 009N, 00AL, R904, 00A1, 00PN, 050513, 10505-13, 183501, 83501, 264-05, 004L,
- Product
- Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22
- Recalling firm
- TriMed Inc.
- Quantity
- 323
- Distribution
- U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
- Recall date
- June 27, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)