FDAClass IIJune 10, 2020

Xhibit Central recall

Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal…

Why was Xhibit Central recalled?

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model No. 96102 Xhibit Central; Software Versions: 1.3.1, 1.3.2, 1.3.3 and 1.3.4; UDI: 10841522100345

Product
Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal…
Recalling firm
Spacelabs Healthcare, Inc.
Quantity
2197 units
Distribution
Worldwide distribution - US Nationwide distribution including in the 40 states and Puerto Rico, and multiple countries including Canada.
Recall date
June 10, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls