FDAClass IIAugust 9, 2023

Incisive CT Power recall

Incisive CT Power (China)-Computed Tomography X-Ray System Model: 728148

Why was Incisive CT Power recalled?

Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: (01)00884838103467 (21) Serial Numbers: 530456 530338 530360 530489 530370 530363 530389 530369 530481 530350 530464 530500 530436 530301 530501 530372 530466 530368 530476 530358 530294 530421 530435 530480 530458 530342 530432 530419 530460 530502 530349 530475 530482 530431 530335 530378 530494 530492 530470 530345 530461 530374 530384 530388 530357 530404 530329 530373 530409 530462 530420 530452 530307

Product
Incisive CT Power (China)-Computed Tomography X-Ray System Model: 728148
Recalling firm
Philips North America
Quantity
53 units OUS
Distribution
Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam
Recall date
August 9, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)