Cardinal Health Jackson-Pratt Channel Drain 15FR recall
Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An introduction/drainage catheter is a device that is a flexible single or…
Why was Cardinal Health Jackson-Pratt Channel Drain 15FR recalled?
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Codes: 1900956, P1900955
- Product
- Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An introduction/drainage catheter is a device that is a flexible single or…
- Recalling firm
- Degania Silicone, Ltd.
- Quantity
- 640 units
- Distribution
- Nationwide Distribution
- Recall date
- August 21, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)