FDAClass IIJune 27, 2018

Fabius GS MRI Anesthesia Machine recall

Fabius GS MRI Anesthesia Machine; Cat. no. 8607300 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms.…

Why was Fabius GS MRI Anesthesia Machine recalled?

The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

ASLA-0017 ASLA-0018 ASLA-0019 ASLA-0020

Product
Fabius GS MRI Anesthesia Machine; Cat. no. 8607300 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms.…
Recalling firm
Draeger Medical, Inc.
Quantity
4
Distribution
US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
Recall date
June 27, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)