Fabius GS Premium Anesthesia Machine recall
Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery…
Why was Fabius GS Premium Anesthesia Machine recalled?
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
ASKM-0102 ASKM-0220 ASKM-0103 ASKM-0221 ASKM-0124 ASKM-0222 ASKM-0125 ASKM-0223 ASKM-0129 ASKM-0224 ASKM-0132 ASKM-0229 ASKM-0133 ASKN-0061 ASKM-0134 ASKN-0090 ASKM-0135 ASLA-0035 ASKM-0136 ASLA-0046 ASKM-0137 ASLA-0047 ASKM-0138 ASLA-0048 ASKM-0139 ASLA-0049 ASKM-0189 ASLA-0050 ASKM-0190 ASLA-0053 ASKM-0191 ASLA-0070 ASKM-0192 ASLA-0071 ASKM-0193 ASLA-0072 ASKM-0194 ASLA-0073 ASKM-0195 ASLA-0074 ASKM-0198 ASLA-0075 ASKM-0199 ASLA-0129 ASKM-0200 ASLA-0131 ASKM-0201 ASLA-0133 ASKM-0202 ASLA-0140 ASKM-0204 ASLA-0141 ASKM-0205 ASLA-0159 ASKM-0207 ASLA-0199 ASKM-0208 ASLA-0200 ASKM-0215 ASLA-0201 ASKM-0219 ASLA-0218
- Product
- Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery…
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 62
- Distribution
- US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
- Recall date
- June 27, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)