Vanguard 360 Femoral Boss Reamer recall
Vanguard 360 Femoral Boss Reamer, Label: REF 32-488474 VANGUARD 360 1NSTRUMENT FEMORAL BOSS REAMER NON-STERILE, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE…
Why was Vanguard 360 Femoral Boss Reamer recalled?
The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog Number: 32-48847 4 Lot Numbers: 022120, 022180, 162530, 232460, 475500 and 475510. 699240, 793350, 793570 Additional Scope: 065030, 095780, 164290, 235870, 253670, 273630, 308620, 418970, 420030, 421210, 570460, 585260, 619690, 708240, 711430, 726490, 816160, 896810, 929690, 953540, 968950
- Product
- Vanguard 360 Femoral Boss Reamer, Label: REF 32-488474 VANGUARD 360 1NSTRUMENT FEMORAL BOSS REAMER NON-STERILE, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE…
- Recalling firm
- Biomet, Inc.
- Quantity
- 461
- Distribution
- California, IA, FL, MD, UT, SC, AL, IL, CO, MA, KS, GA, WA, NV, AR, NH, OK,Pennsylvania, New York, Texas, Virginia, New Jersey, North Carolina, Kentucky, Michigan, Ohio, Wisconsin, Louisiana, Arizona, Indiana, South Dakota, Missouri & Montana. In addition to Korea, B.V., Japan, Puerto Rico, UK, Chile, New Zealand, Australia, & Canada
- Recall date
- August 27, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)