FDAClass IIJuly 27, 2016

Dimension Vista¿ Homocysteine Flex Reagent Cartridge recall

Dimension Vista¿ Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative…

Why was Dimension Vista¿ Homocysteine Flex Reagent Cartridge recalled?

Reagent lots # 15243MA and 15243MB shows a reduced stability once opened that does not meet the Instructions For Use claim of seven (7) days. A maximum bias of -37% was observed.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Number 15243MA Mfg Date: 08/31/2015 Exp Date: 05/31/2016 Lot Number 15243MB Mfg Datge: 08/31/2015 Exp Date: 05/31/2016

Product
Dimension Vista¿ Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative…
Recalling firm
CSL Behring GmbH
Quantity
898 units
Distribution
Distributed to: AL, AR, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WI, WV
Recall date
July 27, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)