Positron Emission Tomography recall
Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 16 PET/CT, GEMINI GXL 16 PET/CT, GEMINI TF Big Bore PET/CT, Philips…
Why was Positron Emission Tomography recalled?
Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS battery cabinet, entered thermal runaway causing the battery to overheat.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GEMINI TF 16 System serial numbers: 7194, 7186, 7235, 7233 GEMINI GXL 16 System serial number: 4201 GEMINI TF Big Bore System serial numbers: 9022, 9201
- Product
- Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 16 PET/CT, GEMINI GXL 16 PET/CT, GEMINI TF Big Bore PET/CT, Philips…
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Quantity
- 7 units
- Distribution
- Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA.
- Recall date
- August 29, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)