FDAClass IIAugust 3, 2016

GE Inspection Technologies recall

GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for 3D metrology and analysis.

Why was GE Inspection Technologies recalled?

Radiation emissions which exceeded the limit in the performance standard were detected during maintenance of one system. The identified cause was insufficient attachment of shielding in the door. Jostling during delivery was identified as the probable cause of this failure to comply with the performance standard.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

SN - PA2384

Product
GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for 3D metrology and analysis.
Recalling firm
GE Inspection Technologies, LP
Quantity
4
Distribution
US Distribution to the states of : FL, TN, and TX.
Recall date
August 3, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)