Edwards SAPIEN 3 Ultra Transcatheter Heart Valve System recall
Edwards SAPIEN 3 Ultra Transcatheter Heart Valve System (SAPIEN 3 Ultra Delivery System) (20 mm) 9630TF20 UDI:00690103201239
Why was Edwards SAPIEN 3 Ultra Transcatheter Heart Valve System recalled?
The action has been initiated to address reports of burst balloons which have resulted in significant difficulty retrieving the device into the sheath and withdrawing the system from the patient during procedures
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All Lots
- Product
- Edwards SAPIEN 3 Ultra Transcatheter Heart Valve System (SAPIEN 3 Ultra Delivery System) (20 mm) 9630TF20 UDI:00690103201239
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 362 units
- Distribution
- US: VA, CA, GA, MA, MD, NY, OR, WA, MN, TN, AL, KS OUS: United Kingdom, Switzerland, Sweden, Spain, Portugal, Poland, Norway, Netherlands, Italy, Ireland, Germany, Finland, Denmark, Canary Islands, Canada, Austria
- Recall date
- August 28, 2019
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)