FDAClass IIAugust 29, 2012

Oxford Gap Gauge SML 3/4mm. The instrument is used for recall

Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the…

Why was Oxford Gap Gauge SML 3/4mm. The instrument is used for recalled?

A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an increase of 1.0mm in the value of each dimension

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog Number: 32-422801 and Lot Number Identification: ZB111101, ZBll1201, ZB120101, ZB120501, ZB111102, and ZB120201.

Product
Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the…
Recalling firm
Biomet U.K., Ltd.
Quantity
119 units
Distribution
Worldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK.