FDAClass IIJuly 3, 2024

NovaGuide 2 Intelligent Ultrasound recall

NovaGuide 2 Intelligent Ultrasound, REFs: NSC-NVGSYS2 & NSC-NVGSYS2-CA

Why was NovaGuide 2 Intelligent Ultrasound recalled?

Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 00850681007399 (US devices), 00850681007481(Canadian devices)/Serial Numbers: 22123240 22033466 22023447 21041633 21122796 22043986 23053678 23033627 22123241 21102756 21102757 21091951 21122798 21041632 22064209 23033624 23033625 22064212 22064207 22064210 21081932 22064208 21041631 21051651 22113121 21102759 21081931 21081933 21122801 21091952 23063852 22113120 22113122 23053679 22113119 23053681 21102760 21122799 22103019 22113113 23063830 23023429 23023421 23063829 22123270

Product
NovaGuide 2 Intelligent Ultrasound, REFs: NSC-NVGSYS2 & NSC-NVGSYS2-CA
Recalling firm
NovaSignal Corp.
Quantity
45 systems
Distribution
Worldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ, AL, CO, CA, OH, MO, FL, TN, NJ, GA, OK, KY, WA and the country of Canada.
Recall date
July 3, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)