FDAClass IIAugust 20, 2014

Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is recall: Software Issue

Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is intended as an intraoperative image-guided localization system to enable minimally invasive…

Why was Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is recalled?

Brainlab Navigation Software Spine & Trauma 3D 2.0/2.1 offers automatic registration of intraoperatively acquired CT image data sets. To enable automatic registration, the software requires the gantry position of the scanner. The gantry position can either be entered manually or submitted automatically from the CT scanner. If automatic gantry communication is available, the so-called &quot

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Catalog Number: 19061-03

Product
Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is intended as an intraoperative image-guided localization system to enable minimally invasive…
Recalling firm
Brainlab AG
Quantity
17 systems
Distribution
Worldwide distribution: US (nationwide) including states of: NC, NY, OH, WI;and countries of: Austria, China, Germany, Italy, Japan, Malaysia, and Saudi Arabia.
Recall date
August 20, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls