FDAClass IIAugust 20, 2014

Medical Positioning Rapid Response HUT table with Dual recall

Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to…

Why was Medical Positioning Rapid Response HUT table with Dual recalled?

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial number--802722, 802368, 803147, 804007, 804132, 803195, 802286, 802303, 802782, 803112, 803200, 803274, 803455, 803468, 803883, 803961 and 803960.

Product
Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to…
Recalling firm
Medical Positioning Inc
Quantity
17 beds
Distribution
Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.
Recall date
August 20, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls