Medical Positioning Rapid Response HUT table with Dual recall
Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to…
Why was Medical Positioning Rapid Response HUT table with Dual recalled?
Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial number--802722, 802368, 803147, 804007, 804132, 803195, 802286, 802303, 802782, 803112, 803200, 803274, 803455, 803468, 803883, 803961 and 803960.
- Product
- Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to…
- Recalling firm
- Medical Positioning Inc
- Quantity
- 17 beds
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.
- Recall date
- August 20, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)