T8 Removal Head recall
T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in…
Why was T8 Removal Head recalled?
One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
RS12K008
- Product
- T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in…
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 9 units
- Distribution
- USA Nationwide Distribution in the state of GA, AK, WA, and CA.
- Recall date
- September 25, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)