Medline Probe Cover Kits containing sterile Ultrasound recall
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATHETERLESS INSERTION KIT, Model Number CVI4510; b) VANTEX 7FR 3L 16CM CVC BUNDLE, Model Number STCVC01145B; c) VANTEX TRIPLE LUMEN BUNDLE, Model Number STCVC2000;
Why was Medline Probe Cover Kits containing sterile Ultrasound recalled?
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lots within expiry, distributed from Medline between December 2017 - May 2023: a) Model Number CVI4510, UDI/DI (EA) 10653160321194, UDI/DI (CS) 00653160321197; b) Model Number STCVC01145B, UDI/DI (EA) 10653160351924, UDI/DI (CS) 00653160351927; c) Model Number STCVC2000, UDI/DI (EA) 10653160336792, UDI/DI (CS) 00653160336795
- Product
- Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATHETERLESS INSERTION KIT, Model Number CVI4510; b) VANTEX 7FR 3L 16CM CVC BUNDLE, Model Number STCVC01145B; c) VANTEX TRIPLE LUMEN BUNDLE, Model Number STCVC2000;
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 620 units
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
- Recall date
- July 26, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)