FDAClass IIJuly 27, 2016

Toshiba America Medical Systems Angio WorkStation recall: Software Issue

Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System

Why was Toshiba America Medical Systems Angio WorkStation recalled?

During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number - XIDF-AWS801

Product
Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System
Recalling firm
Toshiba American Medical Systems Inc
Quantity
23
Distribution
US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.
Recall date
July 27, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)